Total Organic Carbon Analyzer

The carbon analyzer is used to analyze total carbon (TC), inorganic carbon (IC), total organic carbon (TOC), purgeable organic carbon (POC), and non-purgeable organic carbon (NPOC) in solid or liquid samples.

The system comprises a Shimadzu TOC-5000A TOC analyzer, a Shimadzu ASI-5000A autosampler, and a Shimdazu SSM-5000A solid sample module. The TOC-5000A has ultralow sensitivity and measures from 4 ppb to 4000 ppm of total carbon and 4 ppb to 5000 ppm of inorganic carbon. This instrument meets the needs of ultrapure industries, research facilities, and high salt requirements of marine users. SSM-5000A Solid Sample Module details follow.

  • Analyte: TC, IC, TOC (= TC - IC)
  • Method: TC: Catalytically aided combustion oxidation at 900°C IC: Pre-acidification, oven temperature 250°C
  • Measuring range: TC: 0.1 mg to 30 mg carbon IC: 0.1 mg to 20 mg carbon

Sample Prep and Options

The laboratory that houses this instrument contains the necessary support items for sample preparation, including auto pipettes, a balance, nanopure water, and numerous calibration standards. All work with this instrument and within the associated laboratory must be performed in compliance with EMSL practices and permits.

Individuals may use this instrument independently for their research following the necessary training.

 

User Portal Name: Analytical: Total Organic Carbon Analyzer (TOC)
Instrument ID:
1103
Availability: 10 hours a day, 5 days a week
Tom Wietsma (509) 371-6375 wietsma@pnnl.gov
Concern for the potential global change consequences of increasing atmospheric CO2 has prompted interest in the development of mechanisms to reduce...
Two major mechanisms, (bio)chemical alteration and physicochemical protection, stabilize soil organic carbon (SOC) and thereby control soil carbon...
New techniques in the rapid freezing of cells, so that they are vitrified, and cryo-transmission electron microscopy (cryoTEM) of the frozen hydrated...
Previous work with both class F and class C fly ash indicated minimal leaching from most fly ashes tested. However, the addition of NOx removal...
    • Analytes: TC, IC, TOC, NPOC, and POC.
    • Method: Combustion/non-dispersive infrared gas analysis method
    • Combustion temperature: 680°C
    • Measuring range: 4 ppb to 4000 ppm TOC (to 5000 ppm for IC, to 400 ppm for POC)
    • Sample injection volume: 4 to 2000 μL
    • SSM-5000A SOLID SAMPLE MODULE
    • Analyte: TC, IC, TOC ( = TC - IC)
    • Method: TC: Catalytically aided combustion oxidation at 900°C IC: Pre-acidification, oven temperature 250°C
    • Measuring range: TC: 0.1 mg to 30 mg carbon IC: 0.1 mg to 20 mg carbon
    • Sample amount: 1 g (aqueous content < 0.5 g)